Analog Output signals (~ 11 µApp/1 Vpp)
Prepare the oscilloscope as described below:
Vertical deflection
Switch channels A and B to chop mode (CHOP
).
Set the deflection coefficient (Sensitivity
) of the
channels A and B to 0.5 V/DIV
(11µApp), 0.2 V/DIV
(1Vpp).
Horizontal deflection
Set time coefficient (Time basis
) to 0.5 ms/DIV.
Triggering
Trigger automatically (AUTO
)
Trigger Channel A
Trigger positive edge
Calibration
Switch the input coupling switch (AC/DC/GND) of
the channels A and B to GND
(ç or 0)
Use the Y-position potentiometers of the channels
A and B to shift the electron rays congruently to the
center of the screen
(fig.)
Switch the input coupling switch (AC/DC/GND) of
the channels A and B to DC
.
PWM 8 Settings analog signals
* Interface board 1 Vpp
Possible oscilloscope display x - t or x - y
BNC BNC A Ue1
(A)*
BNC B Ue2
(B)*
MODE
Measure amplitude
[°] 25°