- 40 -
8.4 Specifications of 11µApp Interface Board
Signal amplification (Ie1, Ie2, Ie0): 300
mV
A
Input amplifier:
Maximum signal current: Ie0, Ie1, Ie2: 66 µApp
Maximum frequency at encoder input:
-3 dB: approx. 300 kHz
Note:
The maximum input frequency only stands for the cutoff frequency of current/voltage converter in PWM 8
(signal source: frequency generator). In real operation with encoders the frequency highly depends on the
photocells and their capacitance, as well as on the cable lengths.
Measure current/voltage:
Measuring range current: 0 - 500 mA
Measuring range voltage: 0 - 10 V
Tolerance: ± 3 %
Measure amplitudes:
Measuring range: 2 µApp - 33.3 µApp (0.6 - 10 Vpp)
Measuring frequency: 10 Hz - 50 kHz
Tolerance:
without adjustment via software: ± 5 %
with adjustment via software: ± 3 %
Fault detection signal: Ie1 and Ie2 < 4 µApp
Encoder output:
Output signal: like input signal without U
0
8.5 Specifications of 1Vpp Interface Board
Input amplifier:
Maximum signal voltage: ± 5 Vpp
Maximum frequency at encoder input:
for encoder voltage on interface board: approx. 500 kHz
for signals on BNC sockets (- 3dB): approx. 1 MHz
Note:
Input frequencies over 1 MHz are possible; however, the tolerance of the PHA/TV display can no longer be
guaranteed!
The maximum input frequency only stands for the cutoff frequency of the voltage input of PWM 8 (signal
source: frequency generator). In real operation with encoders the frequency highly depends on the encoder
and on the cable lengths.
Measure current/voltage
Measuring range current: 0 - 500 mA
Measuring range voltage: 0 - 10 V
Tolerance: ± 3 %